1. scope of application
Based on silica gel, surface bondedC18(ODS)、C8(MOS) 、 C4(Butyl) 、 C1(SAS)、C6H5(Phenyl) Reversed-phase chromatography of functional groups (RPC,Reversed Phase Chromatography) A series of chromatographic columns with packing materials.
2. performance acceptance
After receiving the chromatographic column, after unpacking inspection and confirmation, please
Perform the test according to the chromatographic conditions in the column evaluation report
test to confirm column quality.
sample:
① Uracil(5.0× 10-6g/mL)
② Acetophenone(2.0× 10-5g/mL)
③ Toluene(7.5× 10-4g/mL)
④ Ethylbenzene(6.0× 10-4g/mL)
⑤ Fluorene(1.0× 10-5g/mL)
All samples were dissolved and diluted in mobile phase.
Note: If the above conditions are inconsistent with the chromatographic column test report,
Subject to the test report.
3. Column connection
The column inlet and outlet are connected in the direction indicated by the arrows on the column label.
The port part of the chromatographic column connecting pipe must be flat, without burrs and cut slopes. Under the possible conditions, the connecting pipes at both ends of the chromatographic column should be as short as possible, and the inner diameter of the connecting pipe should be as small as possible to prevent the real column efficiency of the chromatographic column not being reflected due to sample diffusion.
4. usepHRange of values ??and water resistance
Conventional Column UsepHThe value range is2.0~8.0, special packing columnpHValues ??use a wider range, such asSupersilSeries is1.5~10, SinoPakSeries is1~12. Conventional chromatographic columns use an aqueous phase no higher than90%, the specially treated packing column has higher water resistance, such asSinoChrom ODS-BPThe highest proportion of water phase used in the chromatographic column can reach95%, AQseries of columns can be used99%water box.
The recommended temperature is not higher than60℃。
When using phosphate buffered saline, at medium or highpHIt is recommended that the column temperature not exceed40℃。Avoid sudden temperature rise during use.
6. Column storage
Overnight or short-term storage: When using buffered solutions or saline mobile phases, a10%~20%Methanol/water or acetonitrile/Water replaces the chromatographic column and the mobile phase in the system, and the concentration of organic solvent in the chromatographic column storage solvent is not lower than20%。
Long-term storage: It can be stored with pure methanol or pure acetonitrile, and the chromatographic column plug should be tightened. Columns must be thoroughly cleaned before storage to avoid residual sample impurities. For columns using mobile phases with higher salt concentrations, prior to storage, they must be10%~20%Methanol/water or acetonitrile/Water replaces the chromatographic column and the mobile phase in the system to avoid salt precipitation and crystallization.
Storage temperature is preferably room temperature.
7. Column regeneration
Methanol:water=90:10(v/v), pure methanol, and as mobile phases, washed in sequence, and each mobile phase flows through the chromatographic column not less than20~30times the column volume. Then , pure methanol, methanol:water=90:10(v/v)Do mobile phase washes in sequence.
The regeneration method of the above chromatographic column is not an absolute standard method. Users can choose a suitable solvent that can dissolve the pollutants in the column to wash the mobile phase in the forward or reverse direction according to their actual experience and purpose. However, no matter what method is used to regenerate the column, it is impossible to restore the column efficiency or other parameters to the new column level.
8. Column maintenance
HPLC columns are consumables, and proper operation and maintenance are essential to prolong column life.
(1) Precautions before using the chromatographic column
Unless otherwise specified, the chromatographic column storage solution produced by our company is the mobile phase shown in the evaluation report. Before use, the user must pay attention to the mutual solubility between the column evaluation mobile phase and the actual mobile phase used for the analysis sample.
(2) mobile phase
The organic solvent used in the mobile phase must be chromatographically pure, and the mobile phase water is preferably ultrapure water or double-distilled water in glassware. Solvent filters must be used before buffer salts are used(0.45μmfilter membrane)filter. The mobile phase should be prepared immediately after use, as the salt solution will breed bacteria or precipitate if left for a long time.
(3) sample
Impurities in the sample are one of the main reasons for the contamination of the chromatographic column and the decline in column efficiency. Complicated samples can be pretreated to remove impurities by appropriate methods. Before injection, they must be used0.22μmor0.45μmMembrane filtration. If the sample is inconvenient to handle, it is recommended to use a guard column. Sometimes the mismatch between the sample solvent and the mobile phase (immiscibility, polarity difference, etc.) will lead to poor chromatographic peak shape, ghost peaks, etc. It is recommended to use mobile phase to dissolve the sample as much as possible.
9.Problems and solutions that may occur during use
HPLC columns are prone to leaks during useThe existing problems are the decrease of column efficiency and the sudden increase of column pressure. possibleThe reasons are as follows:
(1) Contamination of the filter sieve in the head of the chromatographic column;
(2) Column head filler contamination;
(3) The salt in the buffer in the chromatographic column forms crystals and precipitates out;
(4) mobile phasepHThe value is beyond the use range of the chromatographic column, which will destroy or dissolve the structure of the stationary phase.
The solution is as follows:
When it is judged that the sieve plate or packing at the head of the column is contaminated, you can use the mobile phase that can dissolve the pollutants, and wash the column in the opposite direction of the column (about20to30times the column volume, or depending on the specific situation; it is best not to connect the detector at this time), flush the pollutants out of the chromatographic column, and then use it in the direction of the arrow (flow direction) marked on the chromatographic column.
When it is judged that there is salt crystallization:20%Rinse the chromatographic column with an aqueous solution of an organic solvent to dissolve all the salts in the column, and then replace it with a mobile phase containing a high concentration of methanol or other organic solvents.
if it is due topHImproper use of values ??often makes it difficult to restore the column.
When the problem cannot be solved by using the above methods, it is recommended to replace the new column